Teledyne API Ultrafine Particle Monitor - Model 651 | APPENDIX A Firmware Commands |
RIS – Read Instrument Status
RIS reads the instrument status (displayed as 13
RIS
Command | RIS |
|
|
Response | X | X = | |
|
| 1 | = Concentration (#/cm3) |
|
| 2 | = Livetime (%) |
|
| 3 | = Not used in Model 651 |
|
| 4 | = Inlet Pressure (mBar) |
|
| 5 | = Nozzle Pressure (%) |
|
| 6 | = Inlet Flow Mode (0.12, 0.6, 3.0) |
|
| 7 | = Analog Input Voltage (mV) |
|
| 8 | = Pulse Height (mV) |
|
| 9 | = Optics Temp (°C) |
|
| 10 = Growth Tube Temp (°C) | |
|
| 11 = Conditioner Temp (°C) | |
|
| 12 = Water Separator Temp (°C) | |
|
| 13 = Water Reservoir (Filled/Not Filled, 0/1) | |
Example |
|
|
|
Command | RIS | Read Instrument Errors | |
Response | 0.00, | Particle Concentration | |
| 100, | Livetime | |
| blank | Not used in Model 651 | |
| 1002, | Inlet Pressure | |
| 100, | Nozzle Pressure | |
| 0.12, | Inlet Flow Mode | |
| 0.00, | Analog Input Voltage | |
| 0, | Pulse Height | |
| 60.0, | Optics Temp | |
| 60.0, | Growth Tube Temp | |
| 20.0, | Conditioner Temp | |
| 7.0, | Water Separator Temp | |
| 0 | Water Reservoir (Filled) |
RL – Read Laser Current
RL reads laser current in mA.
RL
Command | RL |
|
Response | X | X = 0 - 50 |
Example |
|
|
Command | RL | Read laser current |
Response | 30 | 30 mA |
07506C DCN6727 |