Maintenance, Service, and TroubleshootingTeledyne API Ultrafine Particle Monitor - Model 651

T r o u b l e s h o o t i n g

The STATUS screen displays the status of the operating parts in live-time. The table below provides basic information about some status messages and suggestions for corrective action.

Table 9-2 Troubleshooting

Problem

Cause

Suggestions

Nozzle fault indicated on

Low sample flow.

Verify 0.12 L/min inlet flow as described in the Flow

the Home screen and low

 

Checks section of this chapter. If flow is OK, check for

Nozzle Pressure (<50%)

 

the presence of water in the Pressure Transducer

indicated on the Status

 

sample lines.

screen.

Likely causes:

 

 

There is an obstruction in the

The Sample Orifice is likely clogged or dirty and needs

 

Sample Flow Orifice.

cleaning or replacement.

 

The Sample Flow Filter is wet.

Replace the Sample filter if it appears wet. A wet filter

 

This may result after a flooding

may indicate flooding or poor performance of the Water

 

incident, or result from poor

Separator. The problem may be seen as presence of

 

performance of the Water

water in the tubing immediately upstream of the filter.

 

Separator.

Replace the Sample Filter. Also mark flow direction with

 

 

a marker and remove the filter Screen found

 

 

downstream of the Sample Filter. Use compressed air

 

 

to blow back through the Screen to remove trapped

 

 

material. Replace the Screen in its original orientation

 

 

as indicated by the mark.

 

Water present in the pressure

Poor performance of the Water Separator may result if

 

transducer sample lines.

the instrument is operated outside its temperature and

 

 

humidity specification range.

Nozzle fault indicated on

The pressure over the nozzle is

Verify 0.12 L/min inlet flow as described in the Flow

the Home screen and high

high indicating that the nozzle

Checks section of this chapter.

Nozzle Pressure (>300%)

may be plugged or the path

Disconnect the water bottle and remove the inlet block

indicated on the Status

ahead of the nozzle is

and wick assembly as described earlier in this chapter.

screen.

obstructed. Obstruction may be

If the Nozzle Pressure % remains high, the nozzle is

 

due to an improperly installed,

plugged. The nozzle is not user serviceable, contact

 

twisted wick or the presence of

TSI.

 

excess water.

If the Nozzle Pressure % drops to near 100% after wick

 

 

assembly removal, a restriction in the wick cartridge is

 

 

indicated. Unscrew the wick cartridge from the inlet

 

 

block and the growth tube from the conditioner. Look

 

 

through the wick tube to verify an open path. If there is

 

 

no clear path, (e.g. the wick is twisted causing a

 

 

blockage), refer to the manual section on replacing the

 

 

wick.

 

 

If there appears to be excessive water dripping from the

 

 

instrument after inlet removal, flooding may have

 

 

occurred. Disconnect the water fill bottle at the quick

 

 

disconnect and allow the instrument to run for a few

 

 

hours without the wick cartridge installed. This will dry

 

 

the instrument.

88

07506C DCN6727

Page 90
Image 90
Teledyne 651 operation manual Troubleshooting