Teledyne API Ultrafine Particle Monitor - Model 651 Instrument Operation
07506C DCN6727 47
Status Description
Nozzle Pressure The pressure difference upstream and downstream
of the optics assembly. Should be 100% - a 10%
drop in nozzle pressure indicates a nozzle clog.
Separator Temp Temperature of the Separator in degrees Celsius. A
normal Separator temperature is 7°C.
Cabinet Temp Temperature inside the Model 651 cabinet.
Laser Current The operating current of the laser in mA.
Photodetector Indicates photodetector voltage in mV.
Analog Input Displays the voltage of the analog input.
Flow Constant Represented in particles/cm3. Compensates for any
variations in orifice diameter.
SETUP Screens

HOME | SETUP

Pressing the SETUP button on the home screen takes you to the
SETUP screen where you can set the following operating
parameters:
Parameter Description
SAMPLE TIME Select a sample time for updating the display
graph. Choices are 1 sec, 2 sec, 3 sec, 4 sec, 5
sec, 6 sec, 10 sec, 12 sec, 15 sec, 20 sec, 30
sec, 60 sec.
VACUUM Turn the vacuum valve on/off.
INLET FLOW Set the inlet flow. Choices are 3 L/min (total flow),
0.12 L/min (sample flow), and
0.6 L/min (transport flow).
SET TIME Set the time for the internal, real-time clock used
for data logging purposes.
MORE Takes you to the NETWORK SETUP screen.
NETWORK SET
UP
Set up network connections including
NETWORK, ADDRESS, MASK, and GATEWAY.
MORE Takes you to the ADDITIONAL SETUP screen.
ADDITIONAL
SETUP
Specify the ANALOG OUTPUT and LOGGING
time.
ANALOG
OUTPUT
Set an analog voltage range for the output.
LOGGING Choose intervals for logging data.