Teledyne API Ultrafine Particle Monitor - Model 651 | Instrument Operation |
Status
Nozzle Pressure
Separator Temp
Cabinet Temp
Laser Current
Photodetector
Analog Input
Flow Constant
Description
The pressure difference upstream and downstream of the optics assembly. Should be 100% - a 10% drop in nozzle pressure indicates a nozzle clog.
Temperature of the Separator in degrees Celsius. A normal Separator temperature is 7°C.
Temperature inside the Model 651 cabinet. The operating current of the laser in mA. Indicates photodetector voltage in mV. Displays the voltage of the analog input.
Represented in particles/cm3. Compensates for any variations in orifice diameter.
SETUP Screens
HOME SETUP
Pressing the SETUP button on the home screen takes you to the SETUP screen where you can set the following operating parameters:
Parameter | Description |
SAMPLE TIME | Select a sample time for updating the display |
| graph. Choices are 1 sec, 2 sec, 3 sec, 4 sec, 5 |
| sec, 6 sec, 10 sec, 12 sec, 15 sec, 20 sec, 30 |
| sec, 60 sec. |
VACUUM | Turn the vacuum valve on/off. |
INLET FLOW | Set the inlet flow. Choices are 3 L/min (total flow), |
| 0.12 L/min (sample flow), and |
| 0.6 L/min (transport flow). |
|
|
SET TIME | Set the time for the internal, |
| for data logging purposes. |
MORE | Takes you to the NETWORK SETUP screen. |
|
|
NETWORK SET | Set up network connections including |
UP | NETWORK, ADDRESS, MASK, and GATEWAY. |
MORE | Takes you to the ADDITIONAL SETUP screen. |
|
|
ADDITIONAL | Specify the ANALOG OUTPUT and LOGGING |
SETUP | time. |
ANALOG | Set an analog voltage range for the output. |
OUTPUT |
|
LOGGING | Choose intervals for logging data. |
07506C DCN6727 | 47 |