Teledyne API Ultrafine Particle Monitor - Model 651

Product Overview

PARAMETER

Aerosol flow rate

False Background Counts

False background counts

Aerosol Medium

SPECIFICATION

120 ±12 cm3/min

<0.01 particles/ cm3, one hour average

Aerosol medium

Air only

Environmental Operating Conditions

Ambient temp range

Ambient humidity range

Inlet Pressure Operation

Inlet pressure operation (absolute)

Inlet pressure gauge

Water System

10 to 38°C (50 to 100.4°F)

0 to 90% non-condensing

50 to 110 KPa (0.5 to 1.1 atm)

1 to -5 kPa (-20 inch H20)

Condensing liquid

Water system

Water consumption

Vacuum

Distilled water

External 1 liter bottle for up to 4 weeks of operation.

~250 ml/week

Vacuum

Communications

External vacuum pump not included in instrument accessories

Protocol

Interfaces

RS-232

USB

Ethernet

Data Logging

Data logging

Averaging interval

Outputs

ASCII command set

9-pin, D-Sub connector

Type B connector, USB 2.0 compatible at 12 MB

8-wire RJ-45 jack, 10/100 BASE-T, TCP/IP

USB Flash drive

Data averaging interval of 1-3600s 1,2,4,5,6,10,12,15,20,30 or 60s software provides more avg options.

Digital display

Analog output

Digital output

Calibration

Calibration

Power

Requirements

9-inch QVGA color touch screen with graphical interface. Graph of conc vs. time, concentration, time and total counts, and status

BNC connector, 0 to 10V proportional to concentration, or 0 to 7V in LOG concentration mode.

Data download using USB, RS-232 serial, or Ethernet interface

Recommended annually

100 to 240 VAC, 50/60 HZ, 175 W max

07506C DCN6727

19

Page 21
Image 21
Teledyne 651 operation manual Parameter, Water System