Teledyne API Ultrafine Particle Monitor - Model 651 | Product Overview |
PARAMETER
Aerosol flow rate
False Background Counts
False background counts
Aerosol Medium
SPECIFICATION
120 ±12 cm3/min
<0.01 particles/ cm3, one hour average
Aerosol medium
Air only
Environmental Operating Conditions
Ambient temp range
Ambient humidity range
Inlet Pressure Operation
Inlet pressure operation (absolute)
Inlet pressure gauge
Water System
10 to 38°C (50 to 100.4°F)
0 to 90%
50 to 110 KPa (0.5 to 1.1 atm)
1 to
Condensing liquid
Water system
Water consumption
Vacuum
Distilled water
External 1 liter bottle for up to 4 weeks of operation.
~250 ml/week
Vacuum
Communications
External vacuum pump not included in instrument accessories
Protocol
Interfaces
USB
Ethernet
Data Logging
Data logging
Averaging interval
Outputs
ASCII command set
Type B connector, USB 2.0 compatible at 12 MB
USB Flash drive
Data averaging interval of
Digital display
Analog output
Digital output
Calibration
Calibration
Power
Requirements
BNC connector, 0 to 10V proportional to concentration, or 0 to 7V in LOG concentration mode.
Data download using USB,
Recommended annually
100 to 240 VAC, 50/60 HZ, 175 W max
07506C DCN6727 | 19 |