Maintenance, Service, and Troubleshooting

I n s p e c t i n g a n d C l e a n i n g t h e F a n s

The fans should not require much maintenance, but it is beneficial to perform a visual inspection at intervals to check for dust build up. If any of the fans are dusty, blow them clean with compressed air.

C l e a n / R e p l a c e t h e O r i f i c e s

The Model 651 has three orifices:

Sample flow orifice (0.12 L/min)

Transport flow orifice (0.48 L/min)

Auxiliary flow orifice (3 L/min)

Manifold Block

Sample Flow

Orifice (Red)Transport Flow

Orifice (Purple)

Auxiliary Flow

Orifice (Black)

Figure 9-9

Cleaning/Replacing Orifices

To clean or replace an orifice, follow these instructions:

1.Turn off the power to the Model 651.

2.Remove the instrument cover.