Teledyne API Ultrafine Particle Monitor - Model 651

APPENDIX A Firmware Commands

SSTART – Starts a New Sample

SSTART is used to start a new sample routine.

SSTART

Command SSTART,x

Response OK

Example

X = 0,1,2,3 0 = Stop

1 = Start data type 1 (not used)

2 = Start data type 2 (not used)

3 = Start new sample

Response issued after parameters changed.

Command

SSTART,0

Stops sample.

Response

OK

 

Command

SSTART,3

Starts new sample.

Response

OK

 

Command

SSTART

 

Response

3

Parameter not changed – current setting

 

 

displayed on record.

 

 

U records are displayed when the SSTART,3

 

 

command is entered. The records are returned

 

 

once per second (see below).

 

U

Record type

X Elapsed time in seconds (integer)

CConcentration in 1/10th second intervals (float) R Raw counts in 1/10th second intervals (integer)

XNo value represented between commas for the Model 651

TLive time in 1/10th seconds (float) D DTC value (float)

P Absolute pressure in millibars (integer)

AN

Analog input

HM

Pulse height mean in millivolts (integer)

HS

Pulse standard deviation in millivolts (integer)

IS

Instrument status (use RIE command to see a

 

list of statuses)

Example U record: UX,C,C,C,C,C,C,C,C,C,C,R,R,R,R,R,R,R,R,R,R,,T,T,T,T,T,T,T,T,T,T,D ,P,AN,HM,HS,IS

U1,2.13e3,1.97e3,2.18e3,2.24e3,2.34e3,2.29e3,2.26e3,2.31e3,

2.35e3,2.17e3,466,430,475,489,510,500,493,505,512,474,,,,,,,,

,,,0.099,0.099,0.099,0.099,0.099,0.099,0.099,0.099,0.099,0.09

9,1.17,990,0.000,1179,421,0

U2…