Teledyne API Ultrafine Particle Monitor - Model 651 | Unpacking and Setting Up the Model 651 |
WATER FILL
Fitting
WATER EXHAUST
Fitting
Figure
Water Fill and Water Exhaust Fittings
Connecting the Water Exhaust Tube
The waste water should pass into a suitable drain such as a floor drain or a vented container. To connect the drain tube:
1.Push the connector on the supplied length of drain tubing into the WATER EXHAUST fitting on the back panel.
2.Place the other end of the drain tube in a vented container or over a floor drain.
Connecting the Aerosol Supply
The Model 651 allows you to conduct aerosol sampling from either the front or the back of the instrument. To run the instrument effectively, you need an external vacuum capable of drawing
6 SLPM at 400 mbar absolute pressure. Sampling options include the following:
∙Ambient sampling using the inlet screen assembly (provided with the monitor) connected to the Model 651 inlet. The inlet screen assembly prevents large matter (such as insects and dirt) from entering the instrument.
Note: If you are sampling from the back of the instrument, you must use the inlet screen assembly and the flow rate must be 3 L/min when using the inlet screen.
07506C DCN6727 | 27 |