APPENDIX A Firmware Commands | Teledyne API Ultrafine Particle Monitor - Model 651 |
S Record (Status)
S records are displayed on the Text tab when you enter the command RRS in the Firmware Command field. They display status information.
| S |
| Status record identifier |
|
|
|
|
| |||
| Pinlet |
| Inlet pressure in millibars |
|
|
|
|
| |||
| Pvac |
| Vacuum pressure in millibars |
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
| |
| Tcond |
| Temperature of the conditioner in °C |
|
|
| |||||
| Tgrowth |
| Growth tube temperature in °C |
|
|
|
| ||||
| Toptics |
| Optics temperature in °C |
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| |
| Tsep |
| Water separator temperature in °C |
|
|
|
| ||||
| Tinlet |
| Inlet temperature in °C |
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
| Example S Record: |
|
|
|
|
|
|
|
| ||
| S,990,282,20.0,59.9,60.0,7.0,23.2 |
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
| ||
Record Type | Pinlet | Pvac | Tcond |
| Tgrowth | Toptics | Tsep | Tinlet |
| ||
S | 990 | 282 | 20.0 |
| 59.9 | 60.0 | 7.0 | 23.2 |
|
07506C DCN6727 |