Teledyne API Ultrafine Particle Monitor - Model 651 | Maintenance, Service, and Troubleshooting |
S t a t u s M e s s a g e s
Status messages display at the top of the Home screen. The messages are described below.
Status Message
Low Water
Warmup
Laser Fault
Inlet Pressure Fault
Vacuum Fault
Nozzle Fault
Absolute Pressure Fault
Optics Temp Fault
Growth Tube Temp Fault
Conditioner Temp Fault
Separator Temp Fault
Pulse Height Fault
Ready
Fault
Indicator Description
Water level is low Instrument is warming up Laser malfunction
Inlet pressure is too high/low Vacuum pressure is too high/low Flow obstruction exists
Barometric pressure is outside the operating range
Optics temperature is out of range Growth Tube temperature is out of range
Conditioner temperature is out of range
Separator temperature is out of range Low pulse height
Unspecified fault (not covered by any of the specific indicators)
Note: The messages are a warning that there is a problem with the instrument, but only one message can display at a time. Check the Status screen for more details about potential problems.
07506C DCN6727 | 87 |