Unpacking and Setting Up the Model 651 Teledyne API Ultrafine Particle Monitor - Model 651
24 07506C DCN6727
Table 2-2 Model 651 Maintenance Kit PN DU0000169
Qty. Part Number Description
1 DU0000150 Static Dissipative Sample Inlet Tubing (3M)
3 DU0000234 Filter Replacements M651
1 DU0000161 Replacement Critical Flow Control Orifice .005 inch
1 DU0000162
Replacement Critical Transport Flow Control Orifice
.0095 inch
1 DU0000163
Replacement Critical Auxiliary Flow Control Orifice
.0225 inch
1 DU0000157 3783 Wick Cartridge
12 DU0000158 Wick 3783: Replacement Wicks
1 DU0000178 Three-foot length of 1/8 inch tubing
Unpacking
Carefully unpack the Model 651 from the shipping container (refer
to Chapter 10 to avoid damage due to Electro-Static Discharge).
Check to ensure there is no damage to the instrument. If any
damage is found, contact the carrier. Use the Packing List in Table
2-1 to verify that there are no missing components.
Save the original shipping container to be used for future shipping.
If anything is missing, TAPI Technical Support by phone or by email:
Phone: 1-800-324-5190 (within the US)
001-858-657-9800 (outside the US)
858-657-9800 (local)
E-mail: sda_techsupport@teledyne.com.
See Chapter 9 for instructions on how to return the instrument
to TAPI, and Chapter 3 for moving/shipping procedures.
Caution – Prevent Damage and Avoid Warranty Invalidation
The Ultrafine Particle Monitor - Model 651 operates using distilled
(<6 ppm) or HPLC water as a working fluid. Do not tip the instrument
more than 10 degrees during normal operation. Perform the procedures
described in Chapter 3 before moving or shipping the instrument.
Do not:
Ship an “undried” instrument.
Transport an “undried” instrument over long distances.
Subject an “undried” instrument to freezing temperatures.
Any of the above actions can result in the flooding of the optical system,
performance degradation, and possible damage to the instrument. Such
neglect is not covered under the manufacturer’s warranty.