Unpacking and Setting Up the Model 651 |
| Teledyne API Ultrafine Particle Monitor - Model 651 | ||
| Table | |||
| Qty. | Part Number | Description |
|
| 1 | DU0000150 | Static Dissipative Sample Inlet Tubing (3M) |
|
| 3 | DU0000234 | Filter Replacements M651 |
|
| 1 | DU0000161 | Replacement Critical Flow Control Orifice .005 inch |
|
| 1 | DU0000162 | Replacement Critical Transport Flow Control Orifice |
|
|
|
| .0095 inch |
|
| 1 | DU0000163 | Replacement Critical Auxiliary Flow Control Orifice |
|
|
|
| .0225 inch |
|
| 1 | DU0000157 | 3783 Wick Cartridge |
|
| 12 | DU0000158 | Wick 3783: Replacement Wicks |
|
| 1 | DU0000178 |
|
U n p a c k i n g
Carefully unpack the Model 651 from the shipping container (refer to Chapter 10 to avoid damage due to
Save the original shipping container to be used for future shipping.
If anything is missing, TAPI Technical Support by phone or by email:
Phone: | |
| |
| |
sda_techsupport@teledyne.com. |
See Chapter 9 for instructions on how to return the instrument to TAPI, and Chapter 3 for moving/shipping procedures.
Caution – Prevent Damage and Avoid Warranty Invalidation
The Ultrafine Particle Monitor - Model 651 operates using distilled (<6 ppm) or HPLC water as a working fluid. Do not tip the instrument more than 10 degrees during normal operation. Perform the procedures described in Chapter 3 before moving or shipping the instrument.
Do not:
∙Ship an “undried” instrument.
∙Transport an “undried” instrument over long distances.
∙Subject an “undried” instrument to freezing temperatures.
Any of the above actions can result in the flooding of the optical system, performance degradation, and possible damage to the instrument. Such neglect is not covered under the manufacturer’s warranty.
24 | 07506C DCN6727 |