APPENDIX A Firmware CommandsTeledyne API Ultrafine Particle Monitor - Model 651
RLL – Read Liquid Level
RLL reads the liquid level.
RLL
Command | RLL |
|
Response | FULL/NOTFULL X | X = ADC reading from 0 to 4095 |
Example |
|
|
Command | RLL | Read Liquid Level |
Response | FULL (2471) | FULL – water level |
|
| ADC - 2471 |
RPA – Read Absolute Pressure Transducer
RPA reads the absolute pressure transducer in mbars.
RPA
Command | RPA |
|
Response | X | X = A floating point number from 150 to 1150 |
Example |
|
|
Command | RPA | Read Absolute Pressure |
Response | 1001 |
|
RPN – Read Nozzle Pressure Transducer
RPN reads the nozzle pressure transducer in percent.
RPN
Command | RPN |
|
Response | X | X = A floating point number from 0 to 2050 |
Example |
|
|
Command | RPN | Read Nozzle Pressure |
Response | 100 |
|
RPV – Read Vacuum Pressure
RPV reads the vacuum pressure transducer in mbars.
RPV
Command | RPV |
|
Response | X | X = A floating point number from 0 to 1150 |
Example |
|
|
Command | RPV | Read Vacuum Pressure |
Response | 408 |
|
07506C DCN6727 |