APPENDIX A Firmware CommandsTeledyne API Ultrafine Particle Monitor - Model 651

RLL – Read Liquid Level

RLL reads the liquid level.

RLL

Command

RLL

 

Response

FULL/NOTFULL X

X = ADC reading from 0 to 4095

Example

 

 

Command

RLL

Read Liquid Level

Response

FULL (2471)

FULL – water level

 

 

ADC - 2471

RPA – Read Absolute Pressure Transducer

RPA reads the absolute pressure transducer in mbars.

RPA

Command

RPA

 

Response

X

X = A floating point number from 150 to 1150

Example

 

 

Command

RPA

Read Absolute Pressure

Response

1001

 

RPN – Read Nozzle Pressure Transducer

RPN reads the nozzle pressure transducer in percent.

RPN

Command

RPN

 

Response

X

X = A floating point number from 0 to 2050

Example

 

 

Command

RPN

Read Nozzle Pressure

Response

100

 

RPV – Read Vacuum Pressure

RPV reads the vacuum pressure transducer in mbars.

RPV

Command

RPV

 

Response

X

X = A floating point number from 0 to 1150

Example

 

 

Command

RPV

Read Vacuum Pressure

Response

408

 

A-8

07506C DCN6727

Page 110
Image 110