Instrument Operation

Teledyne API Ultrafine Particle Monitor - Model 651

Follow the startup advice contained in the Quick Start Guide (shipped with the instrument).

Standard Operation Procedures

Perform these standard procedures every 4 weeks (800 hours):

Replace the wick.

Check the flow using a volumetric flowmeter.

Fill the water bottle with 1 liter of distilled (<6 ppm) or HPLC water. Do not use tap water.

Verify that the inlet pressure is in the correct operating range relative to the ambient pressure:

o Check the inlet pressure value on the status screen, then disconnect the aerosol inlet and check the value again. The pressure drop caused by an inlet restriction should not exceed 250 mbars (25 kPa).

or

o Check inlet pressure on the status screen, then turn the instrument vacuum off and check the pressure again. The pressure drop caused by an inlet restriction should not exceed 250 mbars (25 kPa).

Check the status screen to make sure the parameters are still accurate.

Check the inlet screen and remove any debris collected there.

Check the time and date on the Flash Drive.

Perform these standard procedures annually:

Replace the filters.

Perform a Zero check by placing a HEPA filter on the Model 651 inlet and ensuring that particle concentration is

<0.01 particles/cm3.

Note: Detailed information about these procedures can be found in the Maintenance section of Chapter 9.