1. 2. 4 Adjustment mode (05)

In the adjustment mode 05, the following adjustment items can be corrected, changed, or checked.

*In code No. column, number after hyphen means sub-code.

 

 

 

 

Acceptable

 

Operation

Code

Description/Mode

Default

Contents

procedure

 

 

 

 

Value

 

group

 

 

 

 

 

 

 

 

 

 

 

 

 

104

Scanner (secondary scanning) copy length re-

128

1~255

When the value increases by 1, the

1

 

production ratio adjustment.

 

 

 

reproduction ratio in the secondary

 

 

 

 

 

 

scanning direction (vertical paper

 

 

 

 

 

 

feeding direction) increases by

 

 

 

 

 

 

approx. 0.1522%.

 

105

Scanner (secondary scanning) start position

128

85~171

When the value increases by 1, the

1

 

deviation

 

 

 

image shifts by approx. 0.1213mm

 

 

 

 

 

 

toward the trailing edge of the pa-

 

 

 

 

 

 

per.

 

106

CCD primary scanning start

For regular

180

5~251

When the value increases by 1, the

1

 

position deviation

copy mode

 

 

image shifts by approx. 0.042mm

 

 

 

 

 

 

toward the front side of the paper

 

 

 

 

 

 

(machine).

 

108

 

For whole-area

133

5~251

When you input a value, which is

1

 

 

copy mode

 

 

47steps (equivalent to2mm)

 

 

 

 

 

 

smaller than the set value of [106],

 

 

 

 

 

 

the rear original edge and thefront

 

 

 

 

 

 

copy edge match (0.042mm/step).

 

135

RADF original stop position (single-sided)

8

0~15

Changes the position where the

1

136

RADF original stop position

 

8

0~15

original stops.When the value in-

1

 

(reverse side of two-sided original)

 

 

creases by 1, the original stop po-

 

 

 

 

 

 

sition shifts by 1 mm away from the

 

 

 

 

 

 

original stopper.

 

137

RADF sensor automatic adjustment and

By pressing the START key, WAIT

6

 

EPROM Initialization

 

 

 

is displayed while the automatic

 

 

 

 

 

 

adjustment is performed.

 

 

 

 

 

 

Perform RADF EPROM Initializa-

 

 

 

 

 

 

tion when EPROM, RADF logic

 

 

 

 

 

 

PWA or sensors are replaced.

 

142

RADF 2-in-1 gap adjustment

 

8

0~15

When the value increases by 1, the

1

 

 

 

 

 

gap between two originals extend

 

 

 

 

 

 

by 1 mm.

 

 

 

 

 

 

 

 

FC-22 ADJUSTMENT

1 - 20

January 2000 © TOSHIBA TEC

Page 28
Image 28
Toshiba FC-22 manual Adjustment mode