SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ECAL

COM Object Reference

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ECAL

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.

ECAL

Object type

 

Property

 

 

Syntax

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.ECAL = Data

Description

 

For channel 1 to channel 4, performs the calibration of the specified ports using

 

 

ECAL. This command makes a measurement of the standard and calculates

 

 

calibration coefficient successively. The calibration executed here is the test set

 

 

calibration specified by SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.

 

 

METHod on page 98.

 

 

 

 

 

NOTE

 

If 2 ports you specify are not a port pair, an error occurs. Also, if

 

 

SCPI.SENSe(Ch).MULTiport.STATe on page 107 is off, an error occurs.

 

 

 

 

 

Variable

 

 

 

 

 

 

 

 

 

 

 

 

Data

 

 

 

 

 

 

 

 

Description

Ports for ECal

 

 

 

 

•Data (0)

port 1 of pair port

 

 

 

•Data (1)

port 2 of pair port

 

 

 

 

 

 

Data type

Variant type (Variant)

 

 

 

 

 

 

 

Note

No query

 

 

 

 

 

 

 

For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.

Examples

 

Dim EcalPort(1) As Variant

 

 

EcalPort(0) = 1

 

 

 

EcalPort(1) = 2

 

 

 

SCPI.SENSe(1).CORRection.MULTiport.COLLect.ECAL = EcalPort

Related objects

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. SAVE on page 100

Equivalent key

 

[CAL] - Test Set Cal - Calibration - Port x-y - Ecal