| SCPI Command Reference |
| |
|
|
Syntax | :SENSe{[1]234}:CORRection:MULTiport:SELF:ONCE |
Description | For channel 1 to channel 4, executes self calibration for the port pairs that are used |
| in the channel with the Test Set calibration. |
|
|
NOTE | If necessary test set calibration has not been executed, this command causes an |
| error. Also, |
Query response | No query |
Example of use | 10 OUTPUT 717;":SENS1:CORR:MULT:SELF:ONCE" |
Related commands | |
| |
| |
Equivalent key | [Cal] - Test Set Cal - Self Cal Once |
Chapter 6 | 79 |