:SENS{1-4}:CORR:MULT:SELF:ONCE

 

SCPI Command Reference

 

:SENS{1-4}:CORR:MULT:SELF:ONCE

 

:SENS{1-4}:CORR:MULT:SELF:ONCE

Syntax

:SENSe{[1]234}:CORRection:MULTiport:SELF:ONCE

Description

For channel 1 to channel 4, executes self calibration for the port pairs that are used

 

in the channel with the Test Set calibration.

 

 

NOTE

If necessary test set calibration has not been executed, this command causes an

 

error. Also, if :SENS{1-4}:MULTon page 81 is off, this command causes an error.

Query response

No query

Example of use

10 OUTPUT 717;":SENS1:CORR:MULT:SELF:ONCE"

Related commands

:SENS{1-4}:MULT on page 81

 

:SENS{1-4}:MULT:PORT{1-2}:SEL on page 82

 

:SENS{1-4}:CORR:MULT:SELF on page 78

Equivalent key

[Cal] - Test Set Cal - Self Cal Once