INSERT LOSS R->XX

Service

Automated Performance Tests

INSERT LOSS R->XX

Measures worst-case insertion loss between the test set’s

and

REFLECTION port or TRANSMISSION port and port XX

INSERT LOSS T->XX

when the test set is switched to port XX. Tests the specified

 

performance of “Insertion Loss,” listed in Table 8-2in

 

Chapter 8, “Specifications.”

 

 

 

Steps:2

 

•Set specified frequency for test set.

 

•Perform transmission Enhanced Response calibration with

 

two test cables (using short, open, load and through). The

 

calibration reference plane should be between the two

 

cables.

 

•Measure insertion loss from each

 

port XX to test set’s REFLECTION port or

 

TRANSMISSION port.

 

•If spec is 16 dB, trace must be above 16 dB.

 

 