Step 4.

 

Making Measurements

 

Measurement Procedures

Step 4.

Executing calibration

 

After the measurement of the calibration standards for necessary ports is finished,

 

calculate and save the calibration coefficients.

 

[Cal] - Test Set Cal - Calibration - Done

 

 

NOTE

The calibration data of the 87075C multi-port test set has higher priority over the

 

analyzer’s normal calibration data. Therefore, if Test Set Cal is turned on with

 

normal calibration data existed, the multi-port test set calibration data is used, and

 

the normal calibration data is ignored.

 

 

26

Chapter 2

Page 34
Image 34