accuracy is compromised with this interpolation.

Test Set Cal and SelfCal

Test Set Cal: An Overview

 

The analyzer will interpolate between calibrated measurement points when the

 

frequency span is narrowed. However, you should be aware that best measurement

 

accuracy is compromised with this interpolation.

Figure 3-1

Frequency Range Example

NOTE

It is recommended that you set your stop frequency to 1300 MHz. The performance

 

of the test set is not specified above 1300 MHz.

 

Using this method of calibration will yield accurate measurements with optimum

 

 

throughput. If your accuracy requirements are more important than your throughput

 

requirements, you may want to perform a separate Test Set Cal for each

 

measurement setup, which eliminates the need for the analyzer to use interpolation.

Chapter 3

33