| SCPI Command Reference |
| |
|
| |||
|
|
| ||
Syntax |
| :SENSe{[1]234}:CORRection:MULTiport:COLLect:ECAL <value 1>,<value 2> | ||
Description |
| For channel 1 to channel 4, performs the calibration of the specified ports using | ||
|
| ECal. This command makes a measurement of the standard and calculates | ||
|
| calibration coefficient successively. The calibration executed here is the test set | ||
|
| calibration specified | ||
|
|
|
| |
NOTE |
| If 2 ports you specify are not a port pair, an error occurs. Also, if | ||
|
| |||
Parameters |
|
|
|
|
|
|
|
|
|
|
|
|
| Description |
|
|
|
|
|
|
| value 1 |
| Specifies a port of the test set port pair to be calibrated. |
|
|
|
|
|
|
| value 2 |
| Specifies the other port of the test set port pair to be |
|
|
|
| calibrated. |
|
|
|
|
|
Query response |
| No query |
| |
Example of use |
| 10 OUTPUT 717;":SENS1:CORR:MULT:COLL:ECAL 1,2" | ||
Related commands |
| |||
|
| |||
Equivalent key |
| [Cal] - Test Set Cal - Calibration - Port |
70 | Chapter 6 |