SCPI Command Reference

 

 

 

:SENS{1-4}:CORR:MULT:COLL:ECAL

 

 

:SENS{1-4}:CORR:MULT:COLL:ECAL

Syntax

 

:SENSe{[1]234}:CORRection:MULTiport:COLLect:ECAL <value 1>,<value 2>

Description

 

For channel 1 to channel 4, performs the calibration of the specified ports using

 

 

ECal. This command makes a measurement of the standard and calculates

 

 

calibration coefficient successively. The calibration executed here is the test set

 

 

calibration specified by :SENS{1-4}:CORR:MULT:COLL:METHon page 73.

 

 

 

 

NOTE

 

If 2 ports you specify are not a port pair, an error occurs. Also, if

 

 

:SENS{1-4}:MULTon page 81 is off, an error occurs.

Parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

 

 

 

 

 

 

 

value 1

 

Specifies a port of the test set port pair to be calibrated.

 

 

 

 

 

 

 

value 2

 

Specifies the other port of the test set port pair to be

 

 

 

 

calibrated.

 

 

 

 

 

Query response

 

No query

 

Example of use

 

10 OUTPUT 717;":SENS1:CORR:MULT:COLL:ECAL 1,2"

Related commands

 

:SENS{1-4}:MULT on page 81

 

 

:SENS{1-4}:CORR:MULT:COLL:METH on page 73

Equivalent key

 

[Cal] - Test Set Cal - Calibration - Port x-y - Ecal

70

Chapter 6

Page 78
Image 78
Agilent Technologies 87075C manual SENS1-4CORRMULTCOLLECAL