:SENS{1-4}:CORR:MULT:COLL:SAVE

 

SCPI Command Reference

 

:SENS{1-4}:CORR:MULT:COLL:SAVE

 

:SENS{1-4}:CORR:MULT:COLL:SAVE

Syntax

:SENSe{[1]234}:CORRection:MULTiport:COLLect:SAVE

Description

For channel 1 to channel 4, calculates the calibration coefficients depending on the

 

selected calibration type. Calibration coefficient is calculated for all port pairs that

 

have all necessary measured calibration data for calculation. If none of the port pair

 

has all necessary data for calculation, an error occurs and the command is ignored.

 

The command clears all measured calibration data of the channel. (No query)

 

 

NOTE

The same calibration method must be used for all port pairs to be used in a channel.

 

Therefore, if calibration is executed after changing the calibration method by

 

:SENS{1-4}:CORR:MULT:COLL:METH on page 73, all the calibration

 

coefficient for the different calibration method will be discarded.

 

 

Query response

No query

Example of use

10 OUTPUT 717;":SENS1:CORR:MULT:COLL:SAVE"

Related commands

:SENS{1-4}:CORR:MULT:COLL:METH on page 73

Equivalent key

[Cal] - Test Set Cal - Calibration - Done