Agilent Technologies 87075C manual Test Set Cal and SelfCal, Test Set Cal An Overview

Models: 87075C

1 169
Download 169 pages 62.06 Kb
Page 52
Image 52
Press Test Set Cal.

NOTE

NOTE

Test Set Cal and SelfCal

Test Set Cal: An Overview

Self Cal is set for each measurement channel and executed for a port pair selected for the channel.

SelfCals are not automatically initiated on currently unused ports. A port is never automatically calibrated unless it is selected— even if the time interval has been exceeded.

NOTE

SelfCals are initiated only when a trigger source is internal and a trigger mode is

 

continuous.

 

 

Step 1.

Press [Cal].

Step 2.

Press Test Set Cal.

Step 3.

Press Timer and enter the interval for executing Self Cal in minutes in the entry area

 

in the upper part of the screen.

Step 4.

Press Self Cal to make the auto Self Cal status on.

 

 

NOTE

You can execute a Self Cal on the currently selected channel by pressing [Cal] - Test

 

Set Cal - Self CalOnce.

 

 

44

Chapter 3

Page 52
Image 52
Agilent Technologies 87075C Test Set Cal and SelfCal, Test Set Cal An Overview, Press Test Set Cal, Set Cal - Self CalOnce