Repeat the following steps for each pair of ports to be

 

Service

 

Automated Performance Tests

XTALK XX->YY

Measures crosstalk resulting from worst-case mismatch

 

(shorts) between the selected ports of the test set. Tests the

 

specified performance of “Crosstalk,” listed in Table 8-2in

 

Chapter 8, “Specifications.”

 

 

 

Steps:3

 

•Set specified frequency for test set.

 

•Set measurement parameters: system bandwidth=30 Hz,

 

Average Factor=8, and Power Level=10.

 

•Connect the analyzer to the test set using the test cables.

 

Repeat the following steps for each pair of ports to be

 

tested:

 

•Connect a through cable between ports XX and YY, and

 

perform Response Cal.

 

•Remove the through cable and connect shorts to ports XX

 

and YY.

 

•Measure transmission (S21).

 

•If the specification is 57 dB, the trace must be below −57

 

dB.

 

 

1.For E5061A/62A analyzers only.

2.Direction of measurement does not matter.

3.Crosstalk measurements done on 11 pairs: port 1 to port 2, port 2 to port 3... port 11 to port 12. Pairs that begin with odd numbers are adjacent pairs. The five pairs that begin with even numbers are considered the worst-case nonadjacent pairs. Other nonadjacent pairs are not tested (for example, port 2 to port 4). This test measures the crosstalk within the testset with worst-case DUT port matches.