COM Object Reference

 

 

SCPI.SENSe(Ch).CORRection.MULTiport.SELF. TIMer

 

SCPI.SENSe(Ch).CORRection.MULTiport.SELF.

 

TIMer

 

Object type

Property

 

Syntax

SCPI.SENSe(Ch).CORRection.MULTiport.SELF.TIMer = Value

 

Value = SCPI.SENSe(Ch).CORRection.MULTiport.SELF.TIMer

Description

For channel 1 to channel 4, specifies the interval at which the self calibration of the

 

test set is executed automatically. If

 

SCPI.SENSe(Ch).CORRection.MULTiport.SELF. STATe on page 102 is on,

 

self calibration is executed at intervals specified by this command.

Variable

 

 

 

 

 

 

 

Value

 

 

 

 

Description

Interval for auto run of self calibration.

 

 

 

 

Data type

integer type (integer)

 

 

 

 

Range

1 to 43200

 

 

 

 

Preset value

60

 

 

 

 

Unit

Minute

 

 

 

 

For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.

Examples

Dim Interval As Integer

 

SCPI.SENSe(1).CORRection.MULTiport.SELF.TIMer = 30

 

Interval = SCPI.SENSe(1).CORRection.MULTiport.SELF.TIMer

Related objects

SCPI.SENSe(Ch).CORRection.MULTiport.SELF. ONCE on page 101

 

SCPI.SENSe(Ch).CORRection.MULTiport.SELF. STATe on page 102

Equivalent key

[CAL] - Test Set Cal - Timer

Chapter 7

103

Page 111