Agilent Technologies 87075C manual Test Set Cal and SelfCal, Test Set Cal An Overview

Models: 87075C

1 169
Download 169 pages 62.06 Kb
Page 46
Image 46
Press Port x-y(a port pair for which you want to execute calibration).

Step 3.

Step 4.

NOTE

Step 5.

NOTE

Step 6.

Step 7.

Step 8.

Step 9.

Test Set Cal and SelfCal

Test Set Cal: An Overview

Press Test Set Cal.

Press Calibration.

If Calibration is disabled, the status of the multi-port test set of this channel may be off. See "Procedure of setting ports for the multi-port test set" in the previous chapter.

Press Port x-y(a port pair for which you want to execute calibration).

The multi-port test set calibration is performed for each port pair. If you want to use, for example, ports 1 and 3 for measurement, calibration must be performed for 1-2 and 3-4.

Connect the open standard to the Port x side and press Port x Open to measure calibration data.

Connect the short standard to the Port x side and press Port x Short to measure calibration data.

Connect the load standard to the Port x side and press Port x Load to measure calibration data.

Connect the open standard to the Port y side and press Port y Open to measure calibration data.

38

Chapter 3

Page 46
Image 46
Agilent Technologies 87075C manual Test Set Cal and SelfCal, Test Set Cal An Overview