Step 3.

Step 4.

NOTE

Step 5.

NOTE

Step 6.

Step 7.

Step 8.

Step 9.

Test Set Cal and SelfCal

Test Set Cal: An Overview

Press Test Set Cal.

Press Calibration.

If Calibration is disabled, the status of the multi-port test set of this channel may be off. See "Procedure of setting ports for the multi-port test set" in the previous chapter.

Press Port x-y(a port pair for which you want to execute calibration).

The multi-port test set calibration is performed for each port pair. If you want to use, for example, ports 1 and 3 for measurement, calibration must be performed for 1-2 and 3-4.

Connect the open standard to the Port x side and press Port x Open to measure calibration data.

Connect the short standard to the Port x side and press Port x Short to measure calibration data.

Connect the load standard to the Port x side and press Port x Load to measure calibration data.

Connect the open standard to the Port y side and press Port y Open to measure calibration data.

38

Chapter 3

Page 46
Image 46
Agilent Technologies 87075C manual Press Test Set Cal Press Calibration