Test Set Cal and SelfCal
Test Set Cal: An Overview
Determine the Test Set Ports to Be Used
To save time, calibrate the lowest number of ports necessary for your test setup. For example, if your test setup requires the use of only 6 of the available twelve ports, calibrate for 6 ports; there is no need to spend the time to calibrate all twelve ports.
The selection of the number of ports is necessary, in part, in order to perform the
•Port 1 to Port 2
•Port 3 to Port 4
•Port 5 to Port 6
The analyzer calculates the remainder of the test set port combinations and saves the data as only one calibration file. The test set calibration can be done in any combination of the adjacent port pairs, [1, 2] and [5, 6] for example. This results in significant
Performing the Test Set Cal with a Mechanical Calibration Kit
This section describes how to perform the Test Set Cal with a mechanical calibration kit.
Step 1. Press [Channel Next] or [Channel Prev] to select the channel for which you want to execute
Step 2. Press [Cal].
*Adjacent ports are those which share an internal through calibration stan- dard. Adjacent ports are: [1, 2], [3, 4], [5, 6], [7, 8], [9, 10], [11, 12].
Chapter 3 | 37 |