:SENS{1-4}:CORR:MULT:COLL:METH

 

 

 

 

SCPI Command Reference

 

 

 

 

:SENS{1-4}:CORR:MULT:COLL:METH

 

 

:SENS{1-4}:CORR:MULT:COLL:METH

Syntax

 

:SENSe{[1]234}:CORRection:MULTiport:COLLect:METHod {ERESSOLT2}

 

 

:SENSe{[1]234}:CORRection:MULTiport:COLLect:METHod?

Description

 

For channel 1 to channel 4, selects the test set calibration from enhanced response

 

 

calibration or full 2-port calibration.

 

 

 

 

NOTE

 

Only ERES (enhanced response calibration) can be specified for the T/R model.

Parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

 

 

 

 

 

 

ERES

Specifies the enhanced response calibration for the test set

 

 

 

 

calibration.

 

 

 

 

 

 

SOLT2

Specifies the full 2-port calibration for the test set

 

 

 

 

calibration.

Example of use

 

 

 

 

10

OUTPUT 717;":SENS1:CORR:MULT:COLL:METH SOLT2"

 

20

OUTPUT 717;":SENS1:CORR:MULT:COLL:METH?"

 

30

ENTER 717;A$

 

Related commands

 

:SENS{1-4}:CORR:MULT on page 69

Equivalent key

 

[Cal] - Test Set Cal - Calibration - Cal Type