Agilent Technologies 87075C manual SENS1-4CORRMULTCOLLLOAD, SCPI Command Reference

Models: 87075C

1 169
Download 169 pages 62.06 Kb
Page 80
Image 80
:SENS{1-4}:CORR:MULT:COLL:LOAD

 

 

SCPI Command Reference

 

 

 

:SENS{1-4}:CORR:MULT:COLL:LOAD

 

 

:SENS{1-4}:CORR:MULT:COLL:LOAD

Syntax

 

:SENSe{[1]234}:CORRection:MULTiport:COLLect[:ACQuire]:LOAD <value>

Description

 

For channel 1 to channel 4, measures the calibration data of the load standard of the

 

 

specified port.

 

 

 

 

 

NOTE

 

If :SENS{1-4}:MULTon page 81 is off, an error occur.

Parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

 

 

 

 

 

 

 

value

 

Specifies a port of the test set for which the calibration data

 

 

 

 

of the load standard is measured.

 

 

 

 

 

Query response

 

No query

 

Example of use

 

10 OUTPUT 717;":SENS1:CORR:MULT:COLL:LOAD 1"

Related commands

 

:SENS{1-4}:MULTon page 81

 

 

:SENS{1-4}:CORR:MULT:COLL:OPEN on page 74

 

 

:SENS{1-4}:CORR:MULT:COLL:SHOR on page 76

 

 

:SENS{1-4}:CORR:MULT:COLL:THRU on page 77

 

 

:SENS{1-4}:CORR:MULT:COLL:ISOL on page 71

Equivalent key

 

[Cal] - Test Set Cal - Calibration - Port x-y - Port x Load (or Port y Load)

72

Chapter 6

Page 80
Image 80
Agilent Technologies 87075C manual SENS1-4CORRMULTCOLLLOAD, SCPI Command Reference