SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.LOAD

 

 

COM Object Reference

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.LOAD

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.

 

ACQuire.LOAD

Object type

Property

 

Syntax

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.ACQuire.LOAD = Value

Description

For channel 1 to channel 4, measures load calibration data for the specified port.

Variable

 

 

 

 

 

 

 

Value

 

 

 

 

Description

Specifies a port of the test set for which the calibration data of the load standard is measured.

 

 

 

 

Data type

integer type (integer)

 

 

 

 

Note

No query

 

 

 

 

For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.

Examples

Dim port_num As Integer

 

port_num = 1

 

SCPI.SENSe(1).CORRection.MULTiport.COLLect.ACQuire.LOAD =

 

port_num

 

Related objects

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.OPEN on

 

page 94

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.SHORt on

page 95

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.THRU on

page 96

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.ACQuire.ISOLation on

page 91

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. SAVE on page 100

Equivalent key [CAL] - Test Set Cal - Calibration - Port x-y - Port x Load (or Port y Load)