Step 5. Activate the Test Set

Introduction and Installation

Installation

 

Step 5. Activate the Test Set

 

1. Connect the semi-rigid cables that were shipped with your test set as shown in

 

Figure 1-6. Check Table 1-1to be sure that you are using the correct cables with

 

your system configuration (the cables you use will be different depending on

 

whether you are using a bench system configuration—as shown in Figure

 

1-6—or a rack configuration).

Figure 1-6

Connect the Front Panel RF Cables

NOTE

Steps 2 and 3 must be performed before you can use your network analyzer to

 

control the multiport test set.

 

2.

Turn on the test set.

 

 

3.

Turn on the analyzer and Press [System] -87075 Setup - 87075 Control to

 

 

activate the test set control.

Chapter 1

17

Page 25
Image 25
Agilent Technologies 87075C manual Activate the Test Set, Introduction and Installation, Connect the Front Panel RF Cables