SCPI Command Reference

 

:SENS{1-4}:CORR:MULT:SELF:TIM

 

:SENS{1-4}:CORR:MULT:SELF:TIM

Syntax

:SENSe{[1]234}:CORRection:MULTiport:SELF:TIMer <value>

 

:SENSe{[1]234}:CORRection:MULTiport:SELF:TIMer?

Description

For channel 1 to channel 4, specifies the interval at which the self calibration of the

 

test set is executed automatically. If :SENS{1-4}:CORR:MULT:SELFon page 78

 

is on, self calibration is executed at intervals specified by this command.

Parameters

 

 

 

 

 

 

 

 

 

 

<value>

 

 

 

 

Description

Interval for auto run of self calibration.

 

 

 

 

Range

1 to 43200

 

 

 

 

Preset value

60

 

 

 

 

Unit

Minute

 

 

 

Query response

<value><newline><^END>

Example of use

10

OUTPUT 717;":SENS1:CORR:MULT:SELF:TIM 30"

 

20

OUTPUT 717;":SENS1:CORR:MULT:SELF:TIM?"

 

30

ENTER 717;A

 

Related commands :SENS{1-4}:MULTon page 81

:SENS{1-4}:CORR:MULT:SELF on page 78

Equivalent key

[Cal] - Test Set Cal - Timer

80