Making Measurements
Measurement Procedures
Calibrating the test set
The procedure to calibrate the test set is described below. Note that the calibration procedure described here is a typical one using a mechanical calibration kit. For more detailed information, including procedures using ECal, see Chapter 3.
Step 1. Selecting a mechanical calibration kit
Select a mechanical calibration kit you want to use for calibration. You can select a mechanical calibration kit using one of the following 2 methods.
[Cal] - Cal Kit - Subsequent corresponding softkeys
[Cal] - Test Set Cal - Calibration - Cal Kit - Subsequent corresponding softkeys
Step 2. Selecting Cal Type
Select a calibration method from full
[Cal] - Test Set Cal - Calibration - Cal Type
Step 3. Selecting an adjacent port pair and executing the measurement of the calibration standards
Select an adjacent port pair for which you want to perform test set calibration. Adjacent port pairs are: [1, 2], [3, 4], [5, 6], [7, 8], [9, 10], and [11, 12]. For the 87075C
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
[Cal] - Test Set Cal - Calibration - Port
Chapter 2 | 25 |