Press Return to return to the port selection menu.

Test Set Cal and SelfCal

Test Set Cal: An Overview

Step 10. Connect the short standard to the Port y side and press Port y Short to measure calibration data.

Step 11. Connect the load standard to the Port y side and press Port y Load to measure calibration data.

Step 12. Connect the thru standard between ports x and y side and press Portx- y Thru to measure calibration data.

Step 13.

Press Return to return to the port selection menu.

Step 14.

Repeat Step 5 through Step 13 for necessary port pairs.

 

 

NOTE

When you calibrate the same port pair in multiple channels, you can reduce the

 

number of standard connections by measuring a standard while changing

 

measurement channels. You don’t need to complete the step 5 through 13 before

 

you perform calibration in other channel, thus you can measure standards for Test

 

Set Cal in parallel in multiple channels.

 

 

Step 15.

Press Done to finish the calibration of the multi-port test set. The calibration

 

coefficients are calculated and saved by this operation.