Agilent Technologies 87075C Save, SCPI.SENSeCh.CORRection.MULTiport.COLLect, COM Object Reference

Models: 87075C

1 169
Download 169 pages 62.06 Kb
Page 108
Image 108
SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. SAVE

 

COM Object Reference

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. SAVE

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.

 

SAVE

Object type

Property

Syntax

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect.SAVE

Description

For the active trace of channels 1 to 4 (Ch), calculates the calibration coefficients

 

depending on the selected calibration type. Calibration coefficient is calculated for

 

all port pairs that have all necessary measured calibration data for calculation. If

 

none of the port pair has all necessary data for calculation, an error occurs and the

 

command is ignored. The command clears all measured calibration data of the

 

channel. (No query)

 

 

NOTE

The same calibration method must be used for all port pairs to be used in a channel.

 

Therefore, if calibration is executed after changing the calibration method by

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. METHod on page 98, all

 

the calibration coefficient for the different calibration method will be discarded.

 

 

Variable

For information on the variable (Ch), see Table 7-2, “Variable (Ch),” on page 91.

Examples

SCPI.SENSe(1).CORRection.MULTiport.COLLect.SAVE

Related objects

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.LOAD on

 

page 93

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.OPEN on

 

page 94

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.SHORt on

 

page 95

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. ACQuire.THRU on

 

page 96

 

SCPI.SENSe(Ch).CORRection.MULTiport.COLLect. METHod on page 98

Equivalent key

[CAL] - Test Set Cal - Calibration - Done