Agilent Technologies E1441A user service Psc?, •Coupled command: No, •*RST Condition: none

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 127
Image 127
*PSC?

*PSC

*PSC 0 1 enables/disables power-on status clear. Clears the Status Byte and Stadard Event register enable masks when power is turned on (*PSC 1). When *PSC 0 is in effect, the Status Byte and Standard Event register enable masks are not cleared when power is turned on. The default setting is 1 (enabled to clear).

Note In order for the *PSC setting to remain in effect through a power failure, a *SAV 0 must be executed after the *PSC command.

*PSC?

*PSC? queries the PSC setting. Returns "0" (*PSC 0) or "1" (*PSC 1).

*RCL

*RCL 0 1 2 3 recalls a previously stored instrument configuration. Four memory locations, 0, 1, 2 and 3, are available to store configurations.

Comments You cannot recall the instrument configuration from a memory location that was not previously specified as a storage location. For example, an error is generated if you attempt to recall from memory location "2" but have not saved to that location with the *SAV command. A +810, "State has not been stored" error is generated if nothing is stored in the specified memory location you are trying to recall.

Coupled command: No

Related commands: *SAV, MEMory:STATe:DELete

*RST Condition: none

*RST

*RST resets the Agilent E1441A as follows:

Sets all commands to the power-on/reset state (see the table following these common commands or at the end of Chapter 2).

Aborts all pending operations.

*RST does not affect:

The output queue

The Service Request and Standard Event Status Enable Registers

The enable masks for the Questionable Signal Registers

Calibration data

Comments Coupled command: No

*RST Condition: none

Chapter 3

Agilent E1441A SCPI Command Reference 127

Page 127
Image 127