Recommended Test Equipment

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, use the accuracy requirements shown to select substitute calibration standards.

Table D-1. Recommended Test Equipment

Instrument

Requirements

Recommended Model

Usea

 

 

 

 

50 feedthrough load

50 Ω ± 0.1

 

Q,P,O,T

 

 

 

 

6 1/2 digit Digital

20 VDC ± 0.01%

Agilent 34401A

Q,P,T

Multimeter (DMM)

Integrating ACrms

 

 

 

10 VACrms ± 0.1%

 

 

 

 

 

 

Thermal Voltage Converter

1kHz to 15 MHz

Ballantine 1395A-3

Q,P

(50 termination, 3V input)

 

(or equivalent 3V TVC)

 

or

 

 

 

Power Meter

100 kHz to 15 MHz

Agilent 437A with

 

or

 

Agilent 8482A

 

 

1 VAC rms ± 0.5%

and 20 dB attenuator

 

Wideband ACrms Meter

 

 

1 kHz to 20 MHz

 

 

 

 

 

 

Frequency Meter

1 ppm accuracy

Agilent 53131A

Q,P,T

 

 

 

 

Oscilloscope

100 MHz

Agilent 54602A

T

 

 

 

 

Spectrum Analyzer

Response to 90 MHz

Agilent 3588A

O

 

 

 

 

a.Q = Quick Verification O= Optional Verification Tests

P = Performance Verification Tests T = Troubleshooting

Test Considerations

Verify that you have selected the correct power line voltage prior to attempting any test procedure to ensure proper instrument operation.

For optimum performance, all test procedures should comply with the following recommendations:

Verify the function generator is set to its reset state (*RST).

Assure that the calibration ambient temperature is stable and between 18° C and 28° C.

Assure ambient relative humidity is less than 80%.

Allow a 30 minute warm-up period before verification or adjustment.

Use only RG-58 or equivalent 50 cable.

Keep cables as short as possible, consistent with the impedance requirements.

Appendix D

Service Procedures 165

Page 165
Image 165
Agilent Technologies E1441A user service Recommended Test Equipment, Test Considerations