Agilent Technologies E1441A Frequency-ShiftKeying FSK Commands, FREQuency:STARt?, FREQuency:STOP

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 108
Image 108
FREQuency:STARt?

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

frequency

numeric

10E-03 to 15E+06

Hz

 

 

 

 

 

 

 

 

FREQuency:STARt?

[SOURce:]FREQuency:STARt? [MINMAX] queries the start frequency. Returns a value in hertz.

FREQuency:STOP

[SOURce:]FREQuency:STOP <frequency>MINMAX sets the stop frequency. Select from 10 mHz to 15 MHz (100 kHz for triangle and ramp). The sweep is phase continuous over the full frequency range. The default is 1 kHz. MIN = 10 mHz. MAX

=15 MHz. [ Stored in volatile memory ]

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

frequency

numeric

10E-03 to 15E+06

Hz

 

 

 

 

 

 

 

 

FREQuency:STOP?

[SOURce:]FREQuency:STOP? [MINMAX] queries the stop frequency. Returns a value in hertz.

Frequency-Shift Keying (FSK) Commands

Use the APPLy command or the equivalent FUNC:SHAP, FREQ, VOLT, and VOLT:OFFS commands to configure the carrier waveform. Set the carrier frequency between 100 Hz and 15 MHz (100 kHz maximum for triangle and ramp). The default is 1 kHz.

108 Agilent E1441A SCPI Command Reference

Chapter 3

Page 108
Image 108
Agilent Technologies E1441A Frequency-ShiftKeying FSK Commands, FREQuency:STARt?, FREQuency:STOP?, Parameters