Function Gain and Linearity Adjustment

Function Gain and Linearity Adjustment

The function generator stores six calibration constants related to function gain and linearity. The constants are calculated from the adjustment value entered. If the calibration procedure is aborted before all setup steps have been completed, no calibration constants are stored.

1.Use a DMM to measure the function generator ACrms output voltage for SETUP 02 in the following table. These adjustments use a HIGH Z output termination.

Table D-12. Function Gain and Linearity Adjustment Setups

 

Nominal Output

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

 

 

 

 

 

02

1 kHz

7.07 V rms

Adjustment for sine wave gain.

 

 

 

 

03

1 kHz

5.6 V rms

Adjustment for amplitude linearity.

 

 

 

 

04

100 Hz

5.6 V rms

Adjustment for triangle wave gain.

 

 

 

 

05

100 Hz

5.6 V rms

Adjustment for ramp gain.

 

 

 

 

06

100 Hz

10.0 V rms

Adjustment for square wave gain.

 

 

 

 

07

100 Hz

1.1 Vrms

Adjustment for square wave linearity.

 

 

 

 

2.Send the CAL:VALue <measured Vac> to the function generator.

3.Send the CAL? command to the function generator to initiate the calibration.

4.Repeat steps 1, 2 and 3 for SETUP 03 through 07.

5.Perform the “Function Gain and Linearity Verification” on page 168.

Appendix D

Service Procedures 179