Agilent Technologies Agilent E1441A Frequency Accuracy, Agilent E1441A DC Function Offset

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 193
Image 193
Agilent E1441A Frequency Accuracy

Table D-26. Performance Test Record for the Agilent E1441A (Page 3 of 7)

Agilent E1441A Frequency Accuracy

Function

Output

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

 

 

Sine wave

3.5V @ 1.0 kHz

999.99

Hz

 

1000.01

Hz

0.55e-3

>10:1

 

 

 

 

 

 

 

 

 

Square wave

3.5V @ 1.0 kHz

495

Hz

 

505

Hz

1e-4

>10:1

 

500 Hz burst rate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Agilent E1441A Function Gain and Linearity

Function

Output

 

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

 

 

Sine wave

7.0V @ 1.0 kHz

6.93

Vrms

 

7.07

Vrms

6.5e-3

>10:1

 

 

 

 

 

 

 

 

 

Sine wave

5.7V @ 1.0 kHz

5.643

Vrms

 

5.757

Vrms

5.9e-3

9.7:1

 

 

 

 

 

 

 

 

 

 

Triangle wave

5.7V

@ 100 Hz

5.643

Vrms

 

5.757

Vrms

5.9e-3

9.7:1

 

 

 

 

 

 

 

 

 

 

 

Ramp wave

5.7V

@ 100

Hz

5.643

Vrms

 

5.757

Vrms

5.9e-3

9.7:1

 

 

 

 

 

 

 

 

 

 

 

Square wave

10V

@ 100

Hz

9.9

Vrms

 

10.1

Vrms

8e-3

12.5:1

 

 

 

 

 

 

 

 

 

 

Square wave

8.0 V @ 100

Hz

7.92

Vrms

 

8.08

Vrms

7e-3

11.4:1

 

 

 

 

 

 

 

 

 

 

 

Agilent E1441A DC Function Offset

Function

Output

 

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

 

 

 

DC Volts

10.0

Vdc

9.8

Vdc

 

10.2

Vdc

0.25e-3

>10:1

 

 

 

 

 

 

 

 

 

 

DC Volts

-10.0

Vdc

-10.2

Vdc

 

-9.8

Vdc

0.25e-3

>10:1

 

 

 

 

 

 

 

 

 

 

Appendix D

193

Page 193
Image 193
Agilent Technologies Agilent E1441A Frequency Accuracy, Agilent E1441A Function Gain and Linearity, Appendix D