Amplitude Flatness Verification

Amplitude Flatness Verification

This test verifies the output amplitude flatness specification at selected frequencies. If you use a Temperature Voltage Converter (TVC) (the recommended method)

or a wide band ACrms voltmeter (with a 50 feed through load), perform this procedure as described. If you are using a measurement device that requires a transfer measurement (for example, a power meter), make the transfer in the reference measurement at 100 kHz.

Set the function generator to the first output indicated in the table below and make a reference measurement. Set each function generator output shown in the table below. Measure the output of the function generator. Compare the amplitude level measured to the reference measurement +/- the error shown in the table. This test is a 50 output termination test.

Table D-7. Amplitude Flatness Verification

 

 

Agilent E1441A

 

Measurement

 

 

 

 

 

 

 

 

Function

Out

Ampl

Freq

Nominal

Error

 

 

Terma

 

 

 

 

 

 

 

 

 

 

 

Q

Sine wave

50

3.0 Vrms

1.00 kHz

<reference>

 

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

100.00 kHz

<reference>

± 0.03 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

500.00 kHz

<reference>

± 0.045 Vrms

 

 

 

 

 

 

 

Q

Sine wave

50

3.0 Vrms

1.00 MHz

<reference>

± 0.045 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

3.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

5.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

7.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

9.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

11.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

3.0 Vrms

13.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

Q

Sine wave

50

3.0 Vrms

15.00 MHz

<reference>

± 0.06 Vrms

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

Appendix D

Service Procedures 171