Agilent Technologies E1441A Table, FUNCtion:SHAPe SIN|SQU|TRI|RAMP|NOIS|USER|DC, Chapter

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 21
Image 21
Table 2-1.

.

Table 2-1.

 

 

 

 

 

 

 

 

 

 

 

Sine

Square

Triangle

Ramp

Noise

Arb

 

 

 

 

 

 

 

 

 

 

 

AM Carrier

X

X

X

X

 

X

 

 

 

 

 

 

 

 

 

 

 

AM Modulating Wave

X

X

X

X

X

X

 

 

 

 

 

 

 

 

 

 

 

FM Carrier

X

X

X

X

 

X

 

 

 

 

 

 

 

 

 

 

 

FM Modulating Wave

X

X

X

X

X

X

 

 

 

 

 

 

 

 

 

 

 

FSK Modulation

X

X

X

X

 

X

 

 

 

 

 

 

 

 

 

 

 

Burst Modulation

X

X

X

X

 

X

 

 

 

 

 

 

 

 

 

 

 

Frequency Sweep

X

X

X

X

 

X

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use the following command to select the output function:

FUNCtion:SHAPe SINSQUTRIRAMPNOISUSERDC

You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command. Because the APPLy command also changes duty cycle, modulation type, trigger source, and trigger slope, you must place the APPLy command first in any sequence of configuration commands.

Output Frequency As shown below, the output frequency range depends on the function currently selected. The table shows functions in decending order of the maximum frequency. The default frequency is 1 kHz for all functions.

Table 2-2.

 

 

 

 

 

 

 

 

 

Parameter

Parameter

 

Minimum

Maximum

Default

 

 

Name

Type

Function

Frequency

Frequency

Units

 

 

 

 

 

 

 

 

 

 

frequency

numeric

Sine

100 Hz

15 MHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

Square

100 Hz

15 MHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

Built-In Arbs

100 Hz

5 MHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

Ramp

100 Hz

100 kHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

Triangle

100 Hz

100 kHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Chapter 2

Agilent E1441A Application Information 21