Agilent E1441A Specifications

Waveforms Standard Waveforms:

Sine, Square, Triangle, Ramp, Noise, DC

 

volts, Sine (x)/x, Negative Ramp,

 

Exponential Rise, Exponential fall, Cardiac

Frequency Characteristics

Arbitrary Waveforms:

 

 

Waveform Length:

8 to 16,000 points

Amplitude Resolution:

12 bits (including sign)

Sample Rate:

40 MSa / sec

Non-Volatile Memory:

Four (4) 16k waveforms

Sine:

100

Hz - 15 MHz

Square:

100

Hz - 15 MHz

Triangle:

100

Hz - 100 kHz

Ramp:

100

Hz - 100 kHz

Noise (Gaussian):

10 MHz bandwidth

Arbitrary Waveforms:

 

 

8 to 8,192 points

100

Hz - 5 MHz

8,193 to 12,287 points

100

Hz - 2.5 MHz

12,288 to 16,000 points

100

Hz to 200 kHz

Resolution:

10 Hz or 10 digits

Accuracy:

10 ppm in 90 days

 

20 ppm in 1 year

 

18° C - 28° C

Sinewave Spectral Purity

Temperature Coefficient: Aging:
Harmonic Distortion

DC to 20 kHz:

20 kHz to 100 kHz:

100 kHz to 1 Mhz:

1 Mhz to 15 Mhz:

Total Harmonic Distortion

DC to 20 kHz:

Spurious (non-harmonic)

Output (DC to 1 Mhz):

Output ( 1 Mhz):

<2 ppm/ ° C

<10 ppm/yr

<-70 dBc

<-60 dBc

<-45 dBc

<-35 dBc

<0.04%

<-65 dBc

<-65 dBc + 6 dB/octave

Phase Noise:

< -52 dBc in a 30 kHz band