Agilent Technologies E1441A user service Repeat steps 4, 5 and 6 for SETUP 35 through, Appendix D

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 183
Image 183
Table D-16. AC Amplitude (50 Ohms) Adjustment Setups

Figure D-3. AC Amplitude (50 Ohms) Adjustment Connection

Table D-16. AC Amplitude (50 Ohms) Adjustment Setups

 

Nominal Output+

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Adjustment for:

 

 

 

 

34

1 kHz

3.5 Vrms

0 dB Output Attenuator

 

 

 

 

35

1 kHz

2.8 Vrms

2 dB Output Attenuator

 

 

 

 

36

1 kHz

2.23 Vrms

4 dB Output Attenuator

 

 

 

 

37

1 kHz

1.77 Vrms

6 dB Output Attenuator

 

 

 

 

38

1 kHz

1.41 Vrms

8 dB Output Attenuator

 

 

 

 

39

1 kHz

1.12 Vrms

10 dB Output Attenuator

 

 

 

 

40

1 kHz

.887 Vrms

12 dB Output Attenuator

 

 

 

 

41

1 kHz

.704 Vrms

14 dB Output Attenuator

 

 

 

 

42

1 kHz

.559 Vrms

16 dB Output Attenuator

 

 

 

 

43

1 kHz

.442 Vrms

18 dB Output Attenuator

 

 

 

 

44

1 kHz

.350 Vrms

20 dB Output Attenuator

 

 

 

 

45

1 kHz

.281 Vrms

22 dB Output Attenuator

 

 

 

 

46

1 kHz

.223 Vrms

24 dB Output Attenuator

 

 

 

 

47

1 kHz

.177 Vrms

26 dB Output Attenuator

 

 

 

 

48

1 kHz

.141 Vrms

28 dB Output Attenuator

 

 

 

 

49

1 kHz

.112 Vrms

30 dB Output Attenuator

 

 

 

 

5.Send the CAL:VALue <measured Vac> to the function generator.

6.Send CAL? command to the function generator to initiate calibration.

7.Repeat steps 4, 5 and 6 for SETUP 35 through 49.

8.Perform the “AC Amplitude Verification” on page 169.

Appendix D

Service Procedures 183

Page 183
Image 183
Agilent Technologies E1441A Repeat steps 4, 5 and 6 for SETUP 35 through, Perform the “AC Amplitude Verification” on page