Agilent Technologies E1441A user service Table, FREQuency <frequency>|MINimum|MAXimum, Chapter

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 22
Image 22
FREQuency <frequency>MINimumMAXimum

For arbitrary waveforms that you create and download to memory,

the maximum frequency depends on the number of points specified in the waveform. As shown below, the maximum output frequency decreases as you specify more points in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.

Table 2-3.

 

 

 

 

 

 

Number of Arb Points

Minimum Frequency

Maximum Frequency

 

 

 

 

 

 

 

8 to 8,192 (8k)

100 mHz

5 MHz

 

 

 

 

 

 

 

8,193 to 12,287 (12k)

100 mHz

2.5 MHz

 

 

 

 

 

 

 

12,288 to 16,000

100 mHz

200 kHz

 

 

 

 

 

 

 

 

 

 

 

Possible Conflict with Function Change: The output frequency is automatically adjusted if you select a function whose maximum frequency is less than that of the currently active function. For example, if you output a 1 MHz sine wave and then change the function to triangle wave, the function generator will adjust the output to 100 kHz (the upper limit for triangle waves). A -221, “Settings conflict” error is generated and the frequency is adjusted.

Possible Conflict with Duty Cycle (square wave only): For output frequencies above 5 MHz, the duty cycle is limited to values between 40% and 60% (below 5 MHz, the range is 20% to 80%). The duty cycle is automatically adjusted if you select a frequency that is not valid with the present duty cycle. For example, if you set the duty cycle to 70% and then change the frequency to 8 MHz, the function generator will automatically adjust the duty cycle to 60% (the upper limit for this frequency). A -221, “Settings conflict” error is generated and the duty cycle is adjusted.

Use the following command to set the output frequency:

FREQuency <frequency>MINimumMAXimum

You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command. Because the APPLy command also changes duty cycle, modulation type, trigger source, and trigger slope, you must place the APPLy command first in any sequence of configuration commands.

Output Amplitude As shown below, the output amplitude range depends on the function currently selected and the output termination. The default amplitude is 100 mVpp (into 50 ohms) for all functions.

22 Agilent E1441A Application Information

Chapter 2

Page 22
Image 22
Agilent Technologies Table, FREQuency <frequency>|MINimum|MAXimum, Agilent E1441A Application Information, Chapter