Agilent Technologies E1441A user service kHz to 100 kHz

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 186
Image 186
1 kHz to 100 kHz.

1.Use a broadband ACrms voltmeter to measure the ACrms output voltage of the function generator for SETUP 64 in table D-19 then go to step 3 to enter that value. Go to step 2 if using a power meter.

Table D-19. AC Flatness Adjustment Setup (broadband AC Voltmeter)

 

Nominal

+

 

 

Output

 

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Reference for:

 

 

 

 

64

1 kHz

3.0 V rms

1 kHz flatness DAC gain

 

 

 

 

2.A. If you are using a broadband ACrms voltmeter, proceed to step 3.

B. If you are using a power meter capable of measurements at 1 kHz, use the power meter to measure the function generator output for SETUP 83 in table D-20 then go to step 3 to enter that value. (If your power meter does not measure to 1 kHz, see the transfer measurement procedure below).

Table D-20. AC Flatness Adjustment Setup (power meter)

 

Nominal Out-

+

 

 

put

 

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Reference for:

 

 

 

 

83

1 kHz

3.0 V rms

V rms, dBm

 

 

 

 

Power Meter Transfer Measurement Procedure (> 1 kHz)

If you are using a power meter not capable of measurement to 1 kHz, you can perform the transfer measurement at a different frequency. For example, the Agilent 437A Power Meter with the Agilent 8482A probe and 20 dB attenuator are specified to a low frequency of 100 kHz. To use this measurement device, perform step 1, then use SETUP 65 (instead of SETUP 64) to obtain a 100 kHz output. Measure the output with the power meter, record the measured value, perform SETUP 83 and enter the recorded value from SETUP 64 (not a new measurement). Then, perform step 3 (you will use SETUP 65 twice). This procedure assumes the function generator output is flat from

1 kHz to 100 kHz.

C. If you are using a Thermal Voltage Converter (TVC), use the TVC to measure the function generator output and enter the measure- ment for SETUP 84 in table D-21. (TVC values entered are in mVDC).

Table D-21. AC Flatness Adjustment Setup (thermal converter)

SETUP

FREQUENCY

AMPLITUDE

Reference for:

 

 

 

 

84

1 kHz

3.0 V rms

Thermal Voltage Converter

 

 

 

 

186 Service Procedures

Appendix D