To Phase Lock Using the

To Phase Lock Using the

Counted Burst Mode

(requires Option 001)

In the counted burst mode, you can synchronize phase-lock signals using an external trigger from the faceplate Ext Trig terminal or VXIbus TTLTRG lines. The following statements show you how to synchronize two function generators in the counted burst mode using the Ext Trig terminal (send the commands to both function generators):

Send these commands to both function generators:

APPL:SIN 10E+3, 5.0

Set both to the same frequency

BM:NCYC INF

Set burst count to “INFINITY"

TRIG:SOUR EXT

Set trigger source to external

BM:STAT ON

Enable the BMod LAST

Send this command only to the "left" function generator:

OUTP:TRIG:IMM

source trigger signal from "left"

 

instrument to rught instrument

 

 

 

 

 

 

 

 

Figure 2-13.

To use TTLTRG lines send these commands to both generators:

APPL:SIN 10E+3, 5.0

Set both to the same frequency

BM:NCYC INF

Set burst count to “INFINITY"

OUTP:TTLT4 ON

Enable TTLT4 line to source

TRIG:SOUR TTLT4

Set trigger source toTTLT4 line

BM:STAT ON

Enable the burst mode LAST

Send this command only to the "left" function generator:

OUTP:TRIG:IMM

source trigger signal from "left"

You don’t have to connect between the Ext Trig terminals when using the TTLTRG lines.