:STRing?

The calibration message is stored in non-volatile memory and does not change when power has been off or after a remote interface reset.

Example Enter calibration information to record the next calibration date:

CAL:STR 'Cal 4/4/YY, Due 10/4/YY' Enter a calibration message to record the cal date of April 4 and next cal due date as October 4 (YY = year of due date).

:STRing?

CALibration:STRing? queries the calibration message and returns a quoted string (or a null string " " if nothing is present).

Example Query the calibration message:

CAL:STR?

Query the calibration message.

enter statement

Enter value into computer.

:VALue

CALibration:VALue <cal_value> specifies the value of the known calibration signal used by the calibration procedure. See “General Calibration/Adjustment Procedure” on page 177..

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

cal_value

numeric

See Cal/Adj on page 177

none

 

 

 

 

 

 

 

 

Comment *RST does not affect the calibration value.

Example Enter the known value for the calibration source signal:

CAL:VAL 10.0

Enter calibration value .

:VALue?

CALibration:VALue? queries the present calibration value.

Example Query the calibration value:

CAL:VAL?

Query the calibration value.

enter statement

Enter value into computer.