:SECure:STATe?

Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.

The function generator calibration is secured when shipped from the factory and the security code is set to "HP_E1441".

*RST does not change the state.

Example Set the calibration state to unsecured:

CAL:SEC:STAT OFF,HP_E1441

Unsecure function generator calibration.

:SECure:STATe?

CALibration:SECure:STATe? returns a "1" or "0" to show whether the calibration security state is enabled (1) or disabled (0). The number is sent to the output buffer.

Example Query the calibration security state:

CAL:SEC:STAT?

Query function generator calibration

 

security state

enter statement

Enter value into computer.

:SETup

CALibration:SETup <0 1 2 3 . . . 84> configure the function generator's internal state for each of the calibration steps to be performed.

:SETup?

CALibration:SETup? queries the calibration setup number. Returns a value between 0 and 84.

:STRing

CALibration:STRing <quoted string> allows you to record calibration information about your function generator while CAL:SEC:STAT is OFF. For example, you can store information such as the last calibration date and/or the next calibration due date. The calibration message may contain up to 40 characters. The string is stored in non-volatile memory.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

quoted string

discrete

alphanumeric

none

 

 

 

 

 

 

 

 

Comments The calibration message can contain up to 40 characters.

Calibration security state must be OFF to store a string.

Chapter 3

Agilent E1441A SCPI Command Reference 75

Page 75
Image 75
Agilent Technologies E1441A SECure:STATe?, SETup?, STRing, Parameters, Example Query the calibration security state