AC Amplitude Verification

AC Amplitude Verification

This procedure is used to check the output amplitude calibration of the function generator. Verification checks are performed to check the accuracy of the pre-attenuator and post attenuator. Make sure you have read “Test Considerations” on page 165.

Set the function generator for each output indicated in the table on the next page. Use a DMM to measure the ACrms output voltage of the function generator. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

Table D-5. AC Amplitude Verification (High Z)

 

 

Agilent E1441A

 

Measurement

 

 

 

 

 

 

 

 

Function

Out

Ampl

Freq

Nominal

Error

 

 

Terma

 

 

 

 

Q

Sine wave

HIGH Z

7.0 Vrms

1.00 kHz

7.0 Vrms

± 0.070 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

5.7 Vrms

1.00 kHz

5.7 Vrms

± 0.057 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

5.5 Vrms

1.00 kHz

5.5 Vrms

± 0.055 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

4.4 Vrms

1.00 kHz

4.4 Vrms

± 0.044 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

3.5 Vrms

1.00 kHz

3.5 Vrms

± 0.035 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

2.8 Vrms

1.00 kHz

2.8 Vrms

± 0.028 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

2.2 Vrms

1.00 kHz

2.2 Vrms

± 0.022 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

1.7 Vrms

1.00 kHz

1.7 Vrms

± 0.017 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

1.4 Vrms

1.00 kHz

1.4Vrms

± 0.014 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

1.1 Vrms

1.00 kHz

1.1 Vrms

± 0.011 Vrms

 

 

 

 

 

 

 

Q

Sine wave

HIGH Z

0.88 Vrms

1.00 kHz

0.88 Vrms

± 0.0088 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.70 Vrms

1.00 kHz

0.70 Vrms

± 0.0070 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.55 Vrms

1.00 kHz

0.55 Vrms

± 0.0055 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.44 Vrms

1.00 kHz

0.44 Vrms

± 0.0044 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.35 Vrms

1.00 kHz

0.35 Vrms

± 0.0035 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.28 Vrms

1.00 kHz

0.28 Vrms

± 0.0028 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.22 Vrms

1.00 kHz

0.22 Vrms

± 0.0022 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.17 Vrms

1.00 kHz

0.17 Vrms

± 0.0017 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.14 Vrms

1.00 kHz

0.14 Vrms

± 0.0014 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.11 Vrms

1.00 kHz

0.11 Vrms

± 0.0011 Vrms

 

 

 

 

 

 

 

Q

Sine wave

HIGH Z

0.088 Vrms

1.00 kHz

0.088Vrms

± 0.00088 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.070 Vrms

1.00 kHz

0.070 Vrms

± 0.00070 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.055 Vrms

1.00 kHz

0.055 Vrms

± 0.00055 Vrms

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

0.044 Vrms

1.00 kHz

0.044 Vrms

± 0.00044 Vrms

 

 

 

 

 

 

 

Q

Sine wave

HIGH Z

0.036 Vrms

1.00 kHz

0.036 Vrms

± 0.00036 Vrms

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

Appendix D

Service Procedures 169