Agilent Technologies Optional Performance Verification Tests, Agilent E1441A Distortion

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 197
Image 197
Optional Performance Verification Tests

Optional Performance Verification Tests

Table D-26. Performance Test Record for the Agilent E1441A (Page 7 of 7)

Agilent E1441A Square Wave Duty Cycle

Function

Output

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

Square wave

1V @ 300 Hz

-0.02Vdc

 

0.02Vdc

25e-6

>10:1

 

50% duty cycle

 

 

 

 

 

 

 

 

 

 

 

 

Square wave

1V @ 300 Hz

-0.52Vdc

 

-0.48Vdc

25e-6

>10:1

 

25% duty cycle

 

 

 

 

 

 

 

 

 

 

 

 

Square wave

1V @ 300 Hz

0.48Vdc

 

0.52Vdc

25e-6

>10:1

 

75% duty cycle

 

 

 

 

 

 

 

 

 

 

 

 

Agilent E1441A Distortion

Function

Output

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

Sine wave

1.1V @ 20 kHz

reference

 

reference

NA

Pass/Fail

 

 

 

 

 

 

 

2nd Harmonic

40 kHz

>70dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

3rd Harmonic

60 kHz

>70dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

4th Harmonic

80 kHz

>70dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

5th Harmonic

100 kHz

>70dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

Sine wave

1.1V @ 100 kHz

reference

 

reference

NA

Pass/Fail

 

 

 

 

 

 

 

2nd Harmonic

200 kHz

>60dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

3rd Harmonic

300 kHz

>60dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

4th Harmonic

400 kHz

>60dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

5th Harmonic

500 kHz

>60dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

Sine wave

1.1V @ 1 MHz

reference

 

reference

NA

Pass/Fail

 

 

 

 

 

 

 

2nd Harmonic

2 MHz

>45dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

3rd Harmonic

3 MHz

>45dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

4th Harmonic

4 MHz

>45dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

5th Harmonic

5 MHz

>45dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

Sine wave

1.1V @ 15 MHz

reference

 

reference

NA

Pass/Fail

 

 

 

 

 

 

 

2nd Harmonic

30 MHz

>35dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

3rd Harmonic

45 MHz

>35dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

4th Harmonic

60 MHz

>35dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

5th Harmonic

75 MHz

>35dB

 

NA

NA

Pass/Fail

 

 

 

 

 

 

 

Appendix D

197

Page 197
Image 197
Agilent Technologies Optional Performance Verification Tests, Agilent E1441A Square Wave Duty Cycle, Appendix D