Agilent Technologies E1441A trigger source, The command sequence to configure this mode is

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 34
Image 34
trigger source.

trigger source.

Counted Burst Modulation Output

The command sequence to configure this mode is:34 Agilent E1441A Application Informationburst ends when

count reached

Manual backgroundManual background burst starts on rising edge

Burst Modulation Trigger Signal (INTernal timer shown)

1/(Burst Rate)

Figure 2-3. Counted Burst Mode with INTernal Trigger

The command sequence to configure this mode is:

APPLY:<shape>

<freq>,<ampl>,<offset>

set up wave form

BM:STATE

ON

 

enable burst modulation

BM:SOURce

INTernal

this selects the "counted" mode

TRIG:SOURce

INTernal

trigger from internal trigger

 

 

 

timer

 

BM:NCYCles

<cycle_count>

set the burst count

BM:INTernal:RATE <frequency>

set the burst rep rate

Figure 2-3shows the operation of the counted burst mode with external trigger source.

Counted Burst Modulation Output

Manual backgroundManual backgroundburst ends when count reached

Manual backgroundManual background burst starts on edge set by TRIG:SLOPE

Burst Modulation Trigger Signal (TRIG:SOUR EXT, :SLOPE POS)

Figure 2-4. Counted Burst Mode with EXTernal Trigger

The command sequence to configure this mode is:

APPLY:<shape>

<freq>,<ampl>,<offset>

set up wave form

BM:STATE

ON

 

enable burst modulation

BM:SOURce

INTernal

this selects the "counted" mode

TRIG:SOURce

EXTernalTTLTRG<0-7>BUS

use external trigger

BM:NCYCles

<cycle_count>

set the burst count

34 Agilent E1441A Application Information

Chapter 2

Page 34
Image 34